Features

  • Qualified for various registrations (FCC filing, VCCI registration,DAR, NVLAP, A2LA, and NATA authorizations)
  • Mainly composed by ferrite radio wave absorber (TFA).Wide-ranged radio wave absorbency is available in combinationwith radio wave matching devices.
  • Barrier-free doors allowing for easy transportation of large-sized devices.
  • Standard-equipped storage room for amplifiers for the use in the immunity test.
  • Applicable to large-sized devices (maximum specification: a turntable of 10 m in diameter/capacity of 20 tons).
  • GP-IB controllable attached facilities allowing for automatic instrumentation in a short time.
  • Additional underground pit construction available for installing counter devices especially in the measuring test of communication equipment (optional).
10 Method Anechoic Chamber

10 Method Anechoic Chamber

catalog

Standard Specifications

Type
ALC-10U
ALC-10UL
Shield inside dimensions (m)
21.0Lx13.5Wx9.0H
22.0Lx14.0Wx9.0H
Site attenuation characteristic (dB)
Horizontal/Vertical ±3(at 30MHz to 18GHz)
Field distribution uniformity characteristic (dB)
(IEC61000-4-3 conforming)
100(at 150kHz to 1000MHz)
60(at 1GHz to 18GHz)
Shield effectiveness (dB)
IEEE-STD 299-1997
100(at 150kHz to 1000MHz)
60(at 1GHz to 18GHz)
Insertion loss (dB)
(MIL STD-220A conforming)
100(at 150kHz to 10GHz)
Composit absorber
Ferrite absorber (TFA) + EM wave absorber
Measuring chamber shield inside dimensions (including an amplifier room)
8.0L×5.0W×3.1H
Door
dimensions (m)
Anechoic chamber(1)
Anechoic chamber(2)
Measuring room(3)
2.0W×2.0H(Barrier-free structure)
1.0W×2.0H
1.0W×2.0H
Antenna lift
Lifting range (m)
Lifting speed (sec/m)
1.0 to 4.0
5 to 25(3-step selection, GP-IB control)
Turntable
Dimension (m)
Load capacity (kg)
Rotating speed (rpm)
3.0
1000
1.0 to 4.0
(3-step selection, GP-IB control)
5.0
3000
0.5 to 2.0
(3-step selection, GP-IB control)
Attached facilities
ITV system/interphone
Power supply
Power supply panel/filter panel/insulating transformer/lighting, etc.